Basit öğe kaydını göster

dc.contributor.authorKaradas, Muhammet
dc.contributor.authorHatipoğlu, Ömer
dc.contributor.authorHavva, E. R.
dc.contributor.authorTurumtay, Emine Akyüz
dc.date.accessioned2020-12-19T19:40:58Z
dc.date.available2020-12-19T19:40:58Z
dc.date.issued2018
dc.identifier.citationKaradaş, M., Hatipoğlu, Ö., Havva, E.R. & Turumtay, E.A. (2018). Influence of different light-curing units on monomer elution from bulk fill composites. Journal of Adhesion Science and Technology, 32(23), 2631-2646. https://doi.org/10.1080/01694243.2018.1499320en_US
dc.identifier.issn0169-4243
dc.identifier.issn1568-5616
dc.identifier.urihttps://doi.org/10.1080/01694243.2018.1499320
dc.identifier.urihttps://hdl.handle.net/11436/1700
dc.descriptionHatipoglu, Omer/0000-0002-4628-8551; Karadas, Muhammet/0000-0002-3357-6896en_US
dc.descriptionWOS: 000456292100008en_US
dc.description.abstractObjectives: This in vitro study assessed the effect of different light-curing units on the elution of monomers from bulk fill composites with different thicknesses. Methods: Five bulk fill composites (Filtek Bulk Fill Flowable, SonicFill 2, SDR, Tetric N-Ceram Bulk Fill, and Venus Bulk Fill) and one conventional composite (Filtek Z250) were selected for the study. the cylindrical specimens in thickness of 2 mm or 4 mm were prepared and photopolymerized for 20 s with a light-emitting diode (VALO Cordless) or a halogen (Monitex BlueLuxer) light-curing unit. the specimens in glass vials were covered with a 75% ethanol/water solution. Ethanol/water extraction solutions were collected for high-performance liquid chromatography analysis after 24 h, 3 days, and 7 days. the data were analyzed with repeated measures and three-way ANOVA (p < 0.05). Results: the total monomer amount was significantly influenced by light-curing source used and thickness. the highest levels of Bis-GMA and Bis-EMA were eluted from Tetric N-Ceram BulkFill and Venus Bulk Fill, respectively. SonicFill 2 released the highest level of TEGDMA at 4-mm thickness. the highest levels of UDMA release, from 4-mm-thick Filtek Bulk Fill Flowable, were attained using the halogen unit. Conclusions: Light-curing sources affected the number of monomers released by materials. the amount of eluted monomers declined over time. the increased ratio of released monomers to increased thickness is material dependent. the number of residual monomers is highly associated with the resin ratio and crosslinking network of the composites.en_US
dc.description.sponsorshipRecep Tayyip Erdogan University Research FundRecep Tayyip Erdogan University [2015.53001.111.03.01]en_US
dc.description.sponsorship1 This study was supported by Recep Tayyip Erdogan University Research Fund under Grant (Project code: [2015.53001.111.03.01]).en_US
dc.language.isoengen_US
dc.publisherTaylor & Francis Ltden_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectLight curing uniten_US
dc.subjectBulk fill compositesen_US
dc.subjectThicknessen_US
dc.subjectMonomer releaseen_US
dc.titleInfluence of different light-curing units on monomer elution from bulk fill compositesen_US
dc.typearticleen_US
dc.contributor.departmentRTEÜ, Diş Hekimliği Fakültesi, Klinik Bilimler Bölümüen_US
dc.contributor.institutionauthorKaradas, Muhammet
dc.contributor.institutionauthorHavva, E. R.
dc.contributor.institutionauthorTurumtay, Emine Akyüz
dc.identifier.doi10.1080/01694243.2018.1499320
dc.identifier.volume32en_US
dc.identifier.issue23en_US
dc.identifier.startpage2631en_US
dc.identifier.endpage2646en_US
dc.relation.journalJournal of Adhesion Science and Technologyen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US