Basit öğe kaydını göster

dc.contributor.authorSulukan, Erdinç
dc.contributor.authorGümrükçü, Zeynep
dc.date.accessioned2024-11-25T10:24:10Z
dc.date.available2024-11-25T10:24:10Z
dc.date.issued2024en_US
dc.identifier.citationSulukan, E., & Gümrükçü, Z. (2024). Biomechanical effects of different miniplate use on bone and miniplate systems in multiple mandible fracture: A finite element study. Injury, 55(12), 111983. https://doi.org/10.1016/j.injury.2024.111983en_US
dc.identifier.issn0020-1383
dc.identifier.issn1879-0267
dc.identifier.urihttps://doi.org/10.1016/j.injury.2024.111983
dc.identifier.urihttps://hdl.handle.net/11436/9792
dc.description.abstractPurpose: The aim of this study is to evaluate the biomechanical efficiency of different miniplates in the treatment of multiple mandibula fractures. Material and Method: Mandible, miniplates, and screws were modeled using the Solid Works v2015 (Dassault Syste`mes, France) program, Subsequently two fracture lines were created on the right parasymphysis and angulus mandible. Models were divided into two main groups according to the plates used in the anterior fracture line: group A, 2 piece 4-hole-bar-I plate, and group B, ellipse plate. Each group was divided into five subgroups according to the plates used in the posterior fracture line (I, X, G, 3D, E) and 10 study models were created in total. Under three different biting forces (anterior, right, left), maximum von-Mises stresses seen on miniplates/screws, and Pmax/Pmin stresses seen in the cortically/cancellous bone were analyzed using the Ansys 16.2 software (ANSYS, Inc., USA). Data was visualized using a color distribution scale and interpreted. Results: The highest von-Mises stress, seen in plates, was found in the I plate (353.82 MPa) at the angulus region of model A1. The highest Pmax and Pmin stresses, revealed in cortical bone, were found respectively in model A1 (181.63 MPa) and model B2 (115.01 MPa). The ideal results on plates were seen in models B3 and B5, in which E plates were used in the parasymphysis and G/E plates were used in the angulus region. Conclusion: Grid and ellipse plates provide successful results and plate geometry is more critical than number in terms of stress distribution.en_US
dc.language.isoengen_US
dc.publisherElsevieren_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectFinite element methoden_US
dc.subjectFixationen_US
dc.subjectMiniplateen_US
dc.subjectMultiple mandible fractureen_US
dc.subjectOpen reductionen_US
dc.titleBiomechanical effects of different miniplate use on bone and miniplate systems in multiple mandible fracture: A finite element studyen_US
dc.typearticleen_US
dc.contributor.departmentRTEÜ, Diş Hekimliği Fakültesi, Klinik Bilimler Bölümüen_US
dc.contributor.institutionauthorGümrükçü, Zeynep
dc.identifier.doi10.1016/j.injury.2024.111983en_US
dc.identifier.volume55en_US
dc.identifier.issue12en_US
dc.identifier.startpage111983en_US
dc.relation.journalInjuryen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US


Bu öğenin dosyaları:

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster